Technical parameters of the measurement system
Inlet voltage: < ±1000V
Sampling rate: 1.25GS/S
Bandwidth: 100MHz
Number of channels: 2
Resolution: 8 bits
Power: < 1000W
Functional requirements of the control system
a. Motion control
It can automatically control the discharge ball distance to track the charging voltage and display the discharge ball distance value
Control the automatic grounding of the main body
Preset impact times
b. Charging control
The constant current charging method with thyristor voltage regulation is adopted
The charging process of the generator can be manually and automatically controlled. According to the test requirements, the charging voltage and charging time can be adjusted and the charging voltage value can be displayed
When charging in an automatic control mode, it can make the charging voltage rise according to the required charging curve and automatically stabilize at the pre-set charging voltage value
c. Trigger control
Ignition can be triggered manually, automatically or by alarm
It can trigger the ignition of the interceptor with a delay range of 0 to 9.9s
d. Safety interlock control
The entire system is equipped with complete alarm functions such as warning lights and bells for the test area, as well as control interfaces
It has the function of door interlock
It features automatic grounding and safety grounding interlocking with the system, as well as overcurrent and overvoltage protection functions
It has an emergency stop function
4) Functional requirements of the measurement system:
It features waveform display, analysis, plotting and printing functions. It can set measurement parameters and automatically calculate the data of each waveform. The entire system has three basic functions: automatic recording, automatic analysis and report output.