Suitable for testing the arc resistance performance of electrical plastics, resin adhesives, insulation paints, as well as insulation materials such as laminates and copper-clad laminates used in industries such as motors, electrical appliances, and home appliances.
Response within
24 hours
Free warranty
12 month
On-line training
Support
process | Arc current/mA | On-off time/sec | Duration/sec | Test elapsed time/sec |
1/8-10 | 10 | 1/8s pass 7/8s break | 60 | 60 |
1/4-10 | 10 | 1/4s pass 3/4s break | 60 | 120 |
1/2-10 | 10 | 1/2s pass 1/2s break | 60 | 180 |
10 | 10 | Conduction | 60 | 240 |
20 | 20 | Conduction | 60 | 300 |
30 | 30 | Conduction | 60 | 360 |
40 | 40 | Conduction | 60 | 420 |