KD2685 Insulation Resistance Tester
KD2685 Insulation Resistance Tester

KD2685 Insulation Resistance Tester

Applications:

This product is suitable for testing the safety parameters of instruments and other electrical appliances in factories, schools, and research institutions, especially for testing the insulation resistance of capacitors and other media.

  • Response within
    24 hours

  • Free warranty
    12 month

  • On-line training
    Support

Profile:

The KD2685A/86A insulation resistance tester is a measuring instrument for measuring the insulation resistance of media such as capacitors. It works stably and reliably, is easy to operate, and has the function of identifying defects.

This instrument has a high degree of integration, and the internal measurement part mainly includes voltage selection and voltage stabilization; Measurement input and amplification section; Pre set and comparison of non-conforming products; Display section. This instrument has high accuracy, fast speed, and good stability

Features:

Parameter:

1. Test voltage: Users can make special requirements for the specific range.
2. KD2685A 10VDC ± 5%
25VDC    ±5%
50VDC   ±2%
100VDC  ±2%
250VDC  ±2%
500VDC  ±2%
3. KD2686A 25VDC ± 5%
50VDC  ±2%
100VDC ±2%
250VDC  ±2%
500VDC  ±2%
1000VDC ±2%
4. Test scope:
5 × 105~1013 Ω (accuracy within the range of 105~5 × 105 Ω is not assessed)
10V gear: 105~109 Ω
25V gear: 106~1010
50V range: 106~1011 Ω
100V range 106~1011 Ω
250V range: 106~1011 Ω
500V range: 107~1012 Ω
1000V range: 107~1012 Ω (accuracy within the range of 1012~1013 Ω is not assessed)
5. Test accuracy:
Header indication 1.00~2.0 accuracy ± 5% ± 1 grid
Header indication 2.0~5.0 accuracy ± 10% ± 1 grid
Header indication 5.0~10.0 accuracy ± 15%
6. Charging time:
Instrument: Charging time<0.1S
Capacitor: Capacity<1uF, charging time<0.1S
Capacity>1uF, charging time<0.5S
7. Preset range: (0.1~9.9) x voltage selection multiplier x multiplier selection (M Ω).
8. Sorting function: This instrument has a sorting function. The required value is selected by the preset dial switch. When the test value is less than the preset value, the "defective" indicator light will turn on, indicating that it is unqualified. When the test value is greater than the preset value, the "defective" indicator light will turn off, indicating that it is qualified.
9. Header indication selection:
When the setting/measuring button is pressed, the value indicated on the header is the preset value of the dial.
When the setting/measuring button is released, the value indicated on the header is the measured value.
10. Test fixture measurement/discharge selection:
When measuring/discharging is selected as measurement, measure the components on the test fixture. When selected for measurement - the alarm is also in the measurement state, and the alarm function is turned on. If it is not qualified, an alarm can be sounded.
When the measurement/discharge selection is discharge, discharge the components on the test fixture.
11. Temperature: 0 ℃~40 ℃ Humidity: ≤ 80% RH
12. Power Supply Voltage: 220V ± 5% Frequency: 50Hz Power: ≤ 50W
13. Volume and weight: 340 (length) x 350 (width) x 150 (height) mm; Not exceeding 6Kg