Profile:
Measurement display:
Dc charging voltage
Primary current of transformer
Ball gap distance of generator body
Status display:
The cutting state of the main power contactor
Switching status of the grounding device
Trigger state of the generator ball gap
Generator charging voltage polarity state
Control:
The control function has manual, automatic and program control functions, and the functions at each level are relatively independent.
Using thyristor voltage regulation mode, with charging voltage feedback measurement system.
The ignition gap and cut-off gap distance can be manually and automatically adjusted and displayed on the LCD panel.
A clipped trigger pulse with adjustable delay and a feedback system with generator ignition trigger.
The stability of charging voltage can reach 0.5% by using constant current charging mode controlled by function.
The LCD panel can indicate the charging voltage and charging process of the shock generator with an accuracy of 1%.
The charging voltage and charging time can be directly input from the LCD panel.
With charging abnormal protection function, can automatically or manually trigger the ignition pulse
Indication of the working state of the impact generator, such as spontaneous combustion, no trigger, abnormal charging, stable charging, etc.
Grounding and grounding release control of the main body and charging part of the device.
The charging voltage polarity can be automatically changed by the button on the controller
The charging process of the charging voltage can be controlled automatically or manually
The alarm can be automatically or manually sounded
Protection and interlocking:
Overcurrent protection
Overvoltage protection
Charge exception protection
Door switch interlocking
Grounding linkage
Polarity switching linkage
Features:
Easy to adjust the wave, simple operation, good synchronization performance, reliable action
Constant current charging automatic control technology
The loop inductance is small and the load capacity is large
Parameter:
Charging voltage setting range 0 ~ 100kV
Working range 20 ~ 100kV
Setting deviation ≤ ±1%
Instability ≤ ±1%
Setting resolution 0.1kV
Charging time setting range 30~180s
Set resolution 1s
Alarm delay 2s
The setting range of pressure times is 1~999
Protection Setting Overcurrent protection value can be set
Overvoltage protection value can be set
Dynamic charging protection can be set
Trigger pulse greater than 15kV, rise time less than 100ns, polarity can be switched
The output impact stability is greater than 99%